2. Analysis and preventive measures of common faults in the insertion part
(1) Yellow, aging, crystallization and other phenomena occur in the outer skin of the inserted tube, which is mainly caused by the failure to timely wipe off the human mucus and protein on the mirror after the examination of the patient. Crystallization of such substances will make the inserted tube yellow and hard. Preventive measures are mainly to clean the surface of stolen goods in time after inspection, cleaning to soak with enzyme solution, and then clean with water.
(2) There are pinholes, damage and wrinkles in the tube snake tube. The main reasons for this phenomenon are the collision between the tube and sharp objects, the Angle of cleaning the disk ring is too small, the patient's mouth pad falls off, the mirror body is bitten by the patient, and the mirror is clamped when placing. The main prevention measures are to take the mirror, cleaning disk circle Angle should not be too small, beware of collision with sharp objects, be clamped, mirror body bitten by patients.
(3) The guiding light beam inserted inside the tube is dim, yellow or not guiding light, and the guiding image beam appears black spots, mainly because the insertion tube has too large discounted Angle, extrusion, collision, clamping, bitten by the patient, which leads to part of the guiding light beam and the optical fiber of the guiding image beam broken. The preventive measures are the same as in (2).
(4) the pliers pipe is also located in the inserted tube, mainly used for biopsy forceps, foreign body forceps, etc., through the fault is mainly the pliers pipe wall pinhole, crack and other damage, causing the mirror internal water leakage, resulting in the mirror internal electrical short circuit, serious caused CCD burn. This kind of failure is mainly caused by the release of needles, biopsy forceps and foreign body forceps in the pliers pipe when the bending Angle is too large. Preventive measures are: accessories in the end of the bending part of the relaxation through the clamp pipe; Release the needle after the attachment extends out the clamp opening; Close the biopsy forceps before inserting and removing; Use a biopsy forceps with no problems; Leak detection before use.